-
Notifications
You must be signed in to change notification settings - Fork 4
created PD_CALIB_INTENSITY #92
New issue
Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.
By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.
Already on GitHub? Sign in to your account
created PD_CALIB_INTENSITY #92
Conversation
|
I suggest that the |
|
Can do. Should there also be a data item to specifically record this, or is special_details enough? |
|
Great, that looks much more understandable now. Ideally (and I admit this is not always achieved in other places in current dictionaries) it should be possible to automatically use the information in this category to redo the calibration calculations, that is, to rederive the incident intensity / detector response values. I can't see this being possible at the moment as a human would need to decide on the calibration algorithm. Also, it is now clear to me that there are two distinct types of calibrations here: one is for the detector response, and one is for the incident intensity. Both of these calibrations could be present for a given diffractogram. I would suggest therefore two categories, one for incident intensity calibration, and one for detector response calibration. Incident intensity calibration does not depend on |
side ♪: this is what I'm trying to achieve (poorly atm) with the the QPA categories. . I'll start on breaking it up. |
|
OK. I've set up I've also set up The way that I've used this type of calibration before is to collect a standard diffractogram from a NIST SRM, and then immediately collect a diffractogram from the unknown specimen. I then calculate a calibration factor based on the intensity of the SRM pattern (or peak, in my case) based on the (relative) intensity values on the NIST certificate, and apply that to the unknown. Technically, you'd need to do this before each measurement, but you could probably get away with once per day/week, so does the calibration belong to the standard diffractogram, the unknown diffractogram, or even the _diffrn.id? |
|
I agree that
To my mind, a particular intensity calibration always belongs to a particular instrument, so our new |
|
I think I've done as much as I can with this one. |
Yes, the phase and diffractogram ids in this category refer to the phases and diffractogram from which the calibration was calculated, not those to which the calibration applies. |
|
Great to have these in the dictionary. |
This category holds information on the intensity corrections used to calibrate detector_ids.
This is different to monitor counts and variations in incident intensity; this deals with detectors that are physically less/more sensitive that others.