BK_8500

BK Precision 8500 interface library

Command Packet

Byte0 Byte1 Byte2 Byte3-24 Byte25
0xAA Address Command Data CRC

If a CMD packet responds without a 0x12 status, usually when data is being returned, the command ID should be in the Byte3 positions.

Status Packet

Byte0 Byte1 Byte2 Byte3 Byte4-24 Byte25
0xAA Address 0x12 Status Reserved CRC

''' Python self.resp_status_dict = { 0x90: "ERROR: Invalid checksum", 0xA0: "ERROR: Invalid value", 0xB0: "ERROR: Unable to execute", 0xC0: "ERROR: invalid command", 0x80: True, } '''

Op state

Bit Meaning 0 Calculate the new demarcation coefficient 1 Waiting for a trigger signal 2 Remote control state (1 means enabled) 3 Output state (1 means ON) 4 Local key state (0 means not enabled, 1 means enabled) 5 Remote sensing mode (1 means enabled) 6 LOAD ON timer is enabled 7 Reserved

Demand state

Bit Meaning 0 Reversed voltage is at instrument's terminals (1 means yes) 1 Over voltage (1 means yes) 2 Over current (1 means yes) 3 Over power (1 means yes) 4 Over temperature (1 means yes) 5 Not connect remote terminal 6 Constant current 7 Constant voltage 8 Constant power 9 Constant resistance

SCPI

Symbol Response Formats Boolean value, can be 1 or “ON”, 0 or “OFF” Integer value, can be zero, positive or negative integer number Flexible numerical value, can be zero, positive or negative float point numeric value String value, characters enclosed in single or double qNueowtelisne, hex code 0x0Ah Return, hex code 0x0Dh