Releases: xiallc/ProSpect
1.1.12
Change Log
- QuickConfig now sorts detected USB2 devices by hub and port. This allows the detector channels to appear in a stable order corresponding to physical location. In this version the column headers Bus and Slot correspond to USB Hub and Port, respectively.
- Allow preset run times up to the firmware-supported limit. The previous limitation was 2147 seconds, but if the user entered a greater value, an a calculation error would truncate the length as low as 1552 seconds. Now the length is correctly limited to 2147 seconds; much longer times are allowed with supported firmware updates. To confirm extended preset run time support, check the Board Information window for DSP code major version 4, minor version 49 or greater.
- Removed the preset type Preset Energy Livetime, as energy livetime is not implemented in most variants of the product.
- Automatically reselect the Scope tab if it was selected while switching products.
- If an error occurs on startup, a button is available to view the log.
- Display the USB firmware version in Help | About.
- Fixed DSP parameters display after resizing the window.
- [Installer] Offer to download and install Windows 7 SHA-2 cryptography support if it's not already installed and device drivers are to be installed. This system update (released on Windows Update in 2015) is needed for driver verification.
1.1.8
OS Compatibility Notice
Microsoft KB3033929. Windows 7 systems not updated since March 2015 need this update installed manually in order to successfully install the USB2 driver. Fix for Code 52 driver installation error.
Change Log
- [microDXP] Fixed an issue with switching channels stopping active runs.
- [microDXP] Added ADC Offset control in Scope panel.
- [microDXP] Fixed serial port display in QuickConfig and System tab.
- Fixed channel enabled / disabled status setting, this setting is saved to the registry for the xMAP, and saved to config file for STJ boards.
- Fixed gain match exception for failed peak fit.
- Updated Handel documentation release.
1.0.42
OS Compatibility Notice
Windows 7 systems not updated since March 2015 must install Microsoft KB3033929 before installing this version.
Change Log
- Fixed baseline spectrum axis labels.
- Prefix handel version with the release changeset.
- Do not log error if firmware type is not needed.
- [microDXP] Added scalefactor to pulser time to scale to microseconds.
- [microDXP] Added SCA data readout for supported units.
- [microDXP] Fixed an issue with High low variable gain calculation.
1.0.31
1.0.30
OS Compatibility Notice
Windows 7 systems not updated since March 2015 must install Microsoft KB3033929 before installing this version.
Change Log
- Fix for Code 52 driver installation error.
1.0.29
Notice of Breaking Change
Please note that installing ProSpect will update the system's PLX driver from 6.3.1 to 7.1.0. This breaks compatibility with
- xManager 1.2.2 and earlier
- Handel versions prior to v1.2.1
OS Compatibility Notice
Windows 7 systems not updated since March 2015 need Microsoft KB3033929 installed manually in order to successfully install the USB2 driver. Fix for Code 52 driver installation error.
Change Log
- Use currently selected fit method in the ROI table for gain calibration.
- Added detection for PXI slot number in QuickConfig, sort devices by PXI slots.
- Fixed X axis unit parsing when opening saved spectrum file.
- [microDXP] Fixed USB version display in board info window.
- [microDXP] Fixed problem with the dynamic range control.
- [microDXP] Added PCPMB PID to microDXP variant list and updated USB2 driver.
1.0.24
OS Compatibility Notice
Windows 7 systems not updated since March 2015 must install Microsoft KB3033929 before installing this version.
Change Log
- Updated Handel to release v1.2.11.
- [microDXP] Support for 80MHz clock rates.
- Added units to acquisition value headers in exported MCA data.
- Fixed a regression in 1.0.22 generating Mercury4 configurations.
- Replaced the application error dialog with a new one that allows automatically saving logs and screenshots or submitting them to XIA.
- Fixed initialization of combo boxes.
- Updated the USB2 driver to fix an installation verification issue.
1.0.22
USB Support Notice
Install the xia_usb2 3.4.7.3 driver hotfix on top of ProSpect 1.0.22 to fix an error in which Device Manager reports the driver signature cannot be verified. Right click the microDXP device in Device Manager, click Update Driver, and navigate to the unzipped driver package.
Change Log
- Updated Handel to release v1.2.7.
- Detect and connect to microDXP boards with custom Vendor ID.
- Updated the logging system.
- Added trace trigger and SCA support for the microDXP.
- Fixed a UI issue with decimal internationalization.
- Fixed an issue with an unsuccessful startup blocking the serial port.
- Updated installer so that USB2 driver install doesn't require reboot.
- Updated microDXP PARSET and GENSET handling to update fine gain trim when GENSET updates.
- Added support for custom High / Low gain mode.
- Use module_statistics_2 for microDXP.
1.0.16
OS Compatibility Notice
Windows 7 systems not updated since March 2015 must install Microsoft KB3033929 before installing this version.
Release Notes
- Updated Mercury firmware with fix for mapping buffer statistics.
- Replaced ConfWizard with QuickConfig, a single page configuration utility.
- [microDXP] Acquisition values peak_interval and peak_sample are now deprecated, replaced by peakint_offset and peaksam_offset.
- All the helper libraries and device layer libraries with no exported functions (e.g. md.dll, xerxes.dll, fdd.dll, xiafile.dll, xmap.dll, xmap_psl.dll), are now built into a single handel.dll.
0.1.21
Release Notes
- Added USB2 driver installation for 64-bit windows.
- Enabled maxwidth acquisition input control
- Implemented improved fit algorithm in ROI Mean and FWHM calculation. Regular curve fitting and Gaussian fit can be toggled by clicking on the Mean and FWHM column headers on the ROI table.