The mode on the instrument to be set to DC voltage
Actual behaviour
The instrument shows a message on the front panel that the parameter value should be specified as a string. In Python, the commands are executed without exception.
System
Windows 10
If you are using a released version of qcodes (recommended):
0.31.0
The Keithley6500 driver does not set the mode correctly.
Steps to reproduce
Expected behaviour
The mode on the instrument to be set to DC voltage
Actual behaviour
The instrument shows a message on the front panel that the parameter value should be specified as a string. In Python, the commands are executed without exception.
System
Windows 10
If you are using a released version of qcodes (recommended):
0.31.0
Following the manual:
https://download.tek.com/manual/DMM6500-901-01B_Sept_2019_Ref.pdf
the solution is simply to add quotes around the mode value in the command.
Related to #1541
I will add a PR shortly.